Process Polarimeter
for Monitoring Chiral Compounds in Preparative LC and SFC Applications
The POLARmonitor is a high resolution, sensitive polarimetric detector for online monitoring the Optical Rotation of chiral compounds. State-of-the-art microprocessor technology and “no moving parts” mechanical design have resulted in a detector yielding excellent baseline stability and high reproducibility with maximum protection in harsh industrial environment.
Measuring cells, made of stainless steel or other biocompatible materials, are available for a wide span of flows rates and pressure, enabling any applications in LC, SFC and biotechnology process.
Chromatogram e:e
Chromatogram Hydrocarbonate
Technical Data
Principle | Faraday Modulation / Compensation |
Optical system | Monochromatic, High power LED 589 nm |
Measuring range | minimum ± 3° opt. rotation maximum ± 90° opt. rotation (depending on cell length) |
Resolution | 0.00001° (100 mm cell) |
Noise Drift Accurary | 0.00002° opt. rotation 0.00005° /h opt. rotation 0.05% (full scale) |
Detection limit | 5 mg/l [alfa ] = 90° ( direct injection ) |
Absorbtion | Sample max. 99.5 % |
Temperature | Ambient 5 – 45° C Sample 5 – 85° C |
Humidity | Non condensing |
Cell length | 3 – 100 mm |
Flow rate | min. 2 ml/minute max. 80 l/minute |
Pressure | 30 bar 280 bar (SFC) (depending on cell) |
Tubings | 1/16” 1/8” DN15 DN 25 |
Read out (display) | Optical Rotation Specific Rotation Concentration |
Signal out | 1000 mV , ± 1000 mV 4 – 20 mA , 4 –12 -20 mA RS 232c , USB (option) |
Cable length | max. 1200 meter to Controller |
Protection Grade | IP 65 ( Optical unit ) IP 62 ( Controller ) |
Power | 100 - 250 VAC / 50 VA |
Size & weight | 600 x 200 x 120 mm, 24 kg |